Central Control Monitor
4-19
Body Surface Area (BSA)
The BSA is calculated automatically when the height and weight data are entered.
Touch the
Body Surface Area
button to toggle through the selection for the formula
used to calculated the BSA. BSA can be calculated using the Dubois, Boyd, or
Pediatric formulas.
The Dubois formula is BSA = (Height)
.725
x (Weight)
.425
x 0.007184.
The Boyd formula is BSA = 3.207 (Height)
0.3
x (Weight x 1000)
(0.7285 – 0.0188 Log (Weight x 1000))
10000
The Pediatric formula is BSA = (Wt
.5378
x Ht
.3964
) x 0.024265.
Blood Volume
The Blood Volume is automatically calculated using the patient weight information
and is displayed in milliliters. If the patient weight is greater than or equal to 20 kg,
the formula used is 70 mL x Wt (kg). If patient weight is greater than or equal to 10 kg
and less than 20 kg, the formula is 75 mL x Wt (kg). If the patient weight is less than
10 kg, the formula is 80 mL x Wt (kg).
Gas Subtab
From the Gas subtab, the user selects the type of gas to blend with O
2
, sets the low
FiO
2
alarm and calibrates the oxygen analyzer. See the
Electronic Gas Blender
chapter for detailed information about connecting and operating the gas blender.
To Select O
2
Blend:
Touch the
O
2
Blend
button to toggle between Air and 95/5. The O
2
blend selected
(95/5 or Air) will be displayed within the button.
To Set Low FiO
2
Alarm:
1. Touch the
Low FiO
2
Alarm
button. The button changes to white.
2. Use the arrows to change the alarm limit. The large arrows adjust the FiO
2
limit
by increments of 0.5 and small arrows adjust the limit by increments of 0.1.
3. Touch the
Low FiO
2
Alarm
button again to store the new value.
Note:
The low FiO
2
Alarm is retained for the current case only and is reset each time
a Perfusion Screen is loaded.
With air selected as the O
2
blend, the Low FiO
2
Alarm defaults to “---” and has a
range of 0.21 - 0.99. With 95-5 selected as the O
2
blend, the Low FiO
2
Alarm defaults
to “---” and has a range of 0.95 – 0.99. Asterisks are displayed and the alarm is
disabled if the alarm limit is set lower than the default value.
Setup
Gas
02 Blend:
Occluders
Patient
Time/Date
Timers
Pumps
Air
O2 Analyzer:
Calibrate
16:07:51
Low FiO2 Alarm:
Calibration Status:
Not Calibrated
Choices are:
Air
95/5
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...