Central Control Monitor
4-59
Select Pump Type
1. Select the pump type. The default selection is Large Roller Pump. The number
of pumps currently defined and the maximum number of pumps allowed is
displayed. Select pump type before making other selections on this screen.
Select Display Units
If the pump type selected is a roller pump, choose a display unit related to flow rate.
1. Touch the
Default Display Units
button until the desired unit of measure is
displayed. The default selection is L/min.
If the pump type selected is a centrifugal pump, the display units default to RPM. A
centrifugal pump can be configured to display flow rates in L/min if a flowmeter is
added to the configuration and placed on the centrifugal pump icon. (See
Add Flow
Sensors
section for further details.)
Select Pump Color
Available:
Assigned:
No
Selection
Enter New Name >
Or Enter New Name:
Next >
Default Display Units
L/min
To Add Pump:
Cancel
Cardioplegia Pump
No
Current defined:
1
Maximum allowed
6
Arterial
Large Roller Pump
Small Roller Pump
Select Pump Type:
Large Master Roller Pump
Small Master Roller Pump
Centrifugal Pump
Select Pump Name:
Select From List:
ART
CPG
CPG1
CPG2
Vent
Vent1
Backup Pump
Vent2
No
Select Pump Type Menu
Select Display Units Button
choices are:
L/min
RPM
Select Pump Name
Pump Designation
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...