Central Control Monitor
4-46
Priority
Condition
CCM Display Message
Operator Response
Status
O
2
calibration finished.
O2 Analyzer Calibration
Message displays for 10 seconds.
Complete
Warm up in progress.
Gas System Warming Up Message clears when warm up
is
complete.
O
2
sensor exposure
Service O2 Sensor
Backup analyzer should be used until
limit reached.
O
2
sensor is replaced by Terumo
Cardiovascular Systems Service.
Gas system warm up
Calibrate O2 Analyzer
Perform O
2
Analyzer calibration.
complete.
The %O
2
and the FiO
2
Calibrate O2 Analyzer
Inspect gas circuit for kinks, blockages,
readings do not agree.
or leaks. If necessary, use a backup O
2
analyzer.
O
2
Analyzer calibration
Cannot Calibrate O2
Check status code.
failed.
Analyzer XX (may only
Retry calibration. If calibration does
appear in Aux tab)
not pass, use backup until unit can be
12 User pressed cancel
serviced.
button.
13 Flow or FiO
2
knob
turned during calibration
16 O
2
Sensor is expired
17, 19 Blender mechanical
problem
18 Flow mechanical problem
20, 21, 22 Flow meter fault
23 O
2
sensor fault
Flow sensor has become <Flow Sensor Name > :
Reconnect flow sensor. Message
disconnected from the
Disconnected
clears automatically.
module.
CDI 100/101 disconnected No CDI 100 Data
Check CDI 100/101 settings and cable
or not all data being
connections. If message still appears,
detected.
contact Terumo Customer Service.
CDI 500 disconnected
No CDI 500 Data
Check CDI 500 settings and cable
or not all data being
connections. If message still appears,
detected.
contact Terumo Customer Service.
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...