Central Control Monitor
4-40
Priority
Condition
CCM Display Message
Operator Response
Alert
Level sensor alert
CHECK LEVEL:
Fill reservoir above alert level. Start
sensor detects low
< Pump Name >
pump if stopped. Alert clears
level.
< Pump Response >
automatically.
Roller/Centrifugal
< Pump Name >:
Check operation and placement of
pump loss of servo
MONITOR FLOW
servo source. If source is operating
source.
properly, try to initiate servo
operation via the CCM. If message
appears again, discontinue use of
servo operation.
Follower pump cannot
< Follower Pump Name >: Check ratio. Adjust speed of master
maintain M/F ratio.
CANNOT MEET RATIO
pump. If message appears again,
discontinue use of master/follower
operation.
Roller/Centrifugal cannot < Pump Name >:
Check the operation and placement of
maintain servo setpoint.
MONITOR FLOW
the servo source. If source is
operating properly, try to initiate
servo operation via the CCM. If
message appears again, discontinue
use of servo operation.
Primary roller/centrifugal < Primary Pump Name >: Try to restart the pump. If the
pump has detected
OVERCURRENT
message reappears, discontinue use
overcurrent condition.
of the pump until it has been serviced
by Terumo Cardiovascular Systems.
Reassign pump, use standby
equipment or use hand crank.
Primary roller pump
< Primary Pump Name >: Clear the pump raceway or back off the
has detected pump jam
PUMP JAM
occlusion and try to restart the pump.
condition.
If the message reappears,
discontinue use of the pump until it
has been serviced by Terumo
Cardiovascular Systems. Reassign
pump or use standby equipment.
Primary roller pump
< Primary Pump Name >: Try to restart pump. If the message
has detected overspeed
OVERSPEED
reappears, discontinue use of the
condition.
pump until it has been serviced by
Terumo Cardiovascular Systems.
Reassign pump, use standby
equipment or use hand crank.
Roller pump has
< Primary Pump Name > : Stop and start the pump. If the
detected underspeed
UNDERSPEED
message reappears, discontinue use
condition.
of the pump and contact Terumo
Customer Service. Reassign pump,
use standby equipment or use hand
crank.
Roller pump has
< Primary Pump Name > : Stop and start the pump. If the
detected belt slip
BELT SLIP
message reappears, discontinue use
condition.
of the pump and contact Terumo
Customer Service. Reassign pump,
use standby equipment or use hand
crank.
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...