Central Control Monitor
4-69
Add an Occluder
A maximum of 1 occluder is allowed on a Perfusion Screen.
1. Select
Add
.
2. Select Occluder as device type.
3. Name the occluder.
Clamp
Reset
Reset
000:00
Pump
Reset
Reset
000:00
Arterial
Vent
Sucker1
Sucker2
CPG
To Place the Device 'Level',
Touch One of the Valid Location(s) Above:
Do Not Place
Next >
Next >
< Previous
Undo Place
Undo Place
Gas
% O2
FiCO2
0
0
0
Do you want to setup a safety connection for Level?
Event
Response
Yes
No
?
Primary Pump requires Level Detector Alert and Alarm Connections.
Place the Level Detector
The available location for the level detector appears as a white square. Touch the
white square to place the level detector.
Set Safety Connections
After a level detector is placed, the system automatically prompts the user to set a
safety connection for the level detector.
Connections are required between a level detector and the primary pump.
Choose
Yes
to set a safety connection. Choose
No
to defer setting a safety
connection.