Modules
7-16
mm
Hg
mm
Hg
Channel 1
Channel 2
Device Name
Safety
ART Pres
Average
Air
Next
Pressures
Zero
mmHg
Display:
300
360
Alert:
Alarm:
Level
Limits (mmHg)
16:07:51
+ 50
+ 10
– 10
– 50
Choose to display average,
systolic, or diastolic pressure
Pressure Monitoring
4. Attach pressure transducer to perfusion circuit.
Note:
The pressure transducer, and if used, pressure dome, must be located at
the correct elevation recommended by the manufacturer in order to obtain the
correct pressure reading. Use a pressure line isolator if erratic displays occur
during continuous flow because of the particular setup.
Setting Pressure Limits
Default alarm and alert limits are set when a Perfusion Screen is configured.
To change the limits for the current case only:
1. Select the
Pressures
subtab within the Safety tab at the bottom of the CCM or
touch the
Pressure
icon.
2. Ensure the correct device name appears on the device name button.
3. To change the alert or alarm limit, touch the
Alert
or
Alarm
button. The
background of the button will turn white.
4. Use the increase/decrease arrows to adjust the limit.
5. Touch the
Alert
or
Alarm
button again to store the new limit.
6. Close the Safety Screen by touching the
Close
button on the right side of the
screen.
7. Repeat steps 1-6 for each pressure transducer.
The system maintains a 10 mmHg minimum difference between the pressure alert
and pressure alarm limits to ensure that the alert limit is always less than the alarm
limit. Scrolling an alarm setting into an alert setting or vice versa, will cause the value
to blink when the button is pressed.
To disable the alarm limit, set the value higher than 900 mmHg or lower than -250
mmHg. The alarm button will display “
***
” when disabled. Any active physiological
alarm will clear when the alarm is disabled. The alert limit cannot be disabled.
Changes made to alert and alarm limits are only retained during the current perfusion
case. The values will revert to the values set during configuration when the case is
exited.
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...