System Base
3-11
The above message box will be displayed if the battery system has less than one
hour capacity for a fully loaded system
If the Battery Test indicates the batteries have less than one hour capacity AND
the system may not be fully charged, recharge the battery and then repeat the
Battery Test. To fully recharge the battery, the system must be plugged into the
AC (Mains) Power with the AC (Mains) Power Switch turned on. It takes up to 13
hours to fully recharge the batteries, if battery charge is very low.
System Network
The network design allows the different types of system modules and pumps to
be connected to any dedicated receptacle on the base. It also allows multiple
types of modules to be connected to the system. The user can define how these
modules interact with the system and each other and can set alert and alarm
responses and monitoring limits via the Central Control Monitor.
There are 9 module receptacles, 3 pump receptacles, a Central Control Monitor
receptacle and a lamp receptacle on each side. If necessary, a pump can be
plugged into a module receptacle.
Lamp Receptacle
CCM Receptacle
Module Receptacles
Pump Receptacles
WARNING
: If the Battery Capacity Updates indicate less than 60 minutes
battery power AND the system should be fully charged, call Terumo Customer
Service. Do not use the system until the battery system has been serviced.
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...