Electronic Gas Blender
8-13
Error
Gas mixer is not meeting Gas System : Knob Adjust
If not on bypass, calibrate O
2
analyzer.
demand and has disabled Only
If message does not clear or if on
control from the CCM.
bypass, adjust the FiO
2
and gas flow
using the knobs located on the front of
the system. If the knobs cannot be
turned, use standby gas blender.
Do not use gas system until it has been
serviced by Terumo Customer Service.
See the Messages subtab within the Aux
tab for any additional error or status
messages.
Priority Condition
CCM Message
Operator Response
Status Gas flow is < 0.2 L/min.
Gas Flow : Off
Turn gas flow to > 0.2 L/min to clear
message.
O
2
analyzer calibration
Calibrating O2 Analyzer
Message clears when calibration is
in progress.
complete.
O
2
analyzer calibration
O2 Analyzer Calibration Complete Message displays for 10 seconds.
finished
Warm up in progress.
Gas System Warming Up
Message clears when warm up is
complete.
O
2
sensor exposure
Service O2 Sensor
Backup analyzer should be used until
limit reached.
O
2
sensor is replaced by Terumo
Cardiovascular Systems Service.
System just powered up Calibrate O2 Analyzer
Calibrate O
2
analyzer
and is expecting a
calibration.
The %O
2
and the FiO
2
Calibrate O2 Analyzer
Inspect gas circuit for kinks, blockages,
readings do not agree
or leaks. If necessary, use a backup O
2
analyzer.
O
2
Analyzer calibration Cannot Calibrate O2
Check status code.
failed.
Analyzer XX (may only appear Retry calibration. If calibration does
in Message tab)
not pass, use backup O
2
analyzer until
unit can be serviced.
12 User pressed cancel
button.
13 Flow or FiO
2
knob
turned during calibration
16 O
2
Sensor is expired
17, 19 Blender mechanical
problem
18 Flow mechanical problem
20, 21, 22 Flow meter fault
23 O
2
sensor fault
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...