Central Control Monitor
4-45
Priority
Condition
CCM Display Message
Operator Response
Status
Centrifugal pump speed
<Centrifugal Pump Name>: Verify forward flow. Check
reduced to 1550 RPM
Coasting
connected devices. Increase pump
speed when appropriate.
Reconnecting Master
Reconnecting
Message will clear when connection
and Follower
<Master Pump Name> to is completed.
<Follower Pump Name>
Enabling Pulse mode
<Pump Name > :
Pump enters Pulse mode.
Pump Baseline xx%
Pump is ramping to
<Pump Name > :
Ensure pump reaches desired
baseline setting
Ramp Baseline xx%
baseline setting.
Exiting Pulse mode
<Pump Name > :
Message clears after approximately
Leaving Pulse
3 seconds.
Power system detects
Power System at 80%
If possible, reduce power
power usage at 80%.
Capacity
consumption by turning off lamp/s,
and slowing or stopping pumps.
Level system is off,
<Level Detector Name > : Mount the level sensor to reservoir.
level sensor not mounted Not Attached
Message clears automatically.
to the venous reservoir.
Level system is off,
<Level Detector Name > : Reconnect level sensor.
level sensor becomes
Disconnected
Message clears automatically.
disconnected from module.
Level selection is changed <Level Detector Name > : Allow approximately 10 seconds for
from Alert to Alert/Alarm Initializing alarm probe
probe to initialize.
in a Perfusion Screen
Level detection is turned <Level Detector Name > : Message will clear after
OFF during perfusion
Off
approximately 3 seconds.
Pressure alarm set out
<Pressure Sensor Name > : Message will clear when pressure
of range
Alarm Disabled
alarm is set within range or
message is replaced by a higher
priority
message.
Occluder responds to an <Occluder Name > : is
Note occluder value, change if
event.
<open, closed, or % flow> desired.
Uncalibrated occluder
Calibrate <Occluder Name> Calibrate occluder.
detected.
Occluder is disconnected. <Occluder Name > :
Reconnect and calibrate occluder.
Connect Occluder
Gas system flow is
Gas Flow : Off
Turn gas flow to
≥
0.2 L/min to clear
< 0.2 L/min.
message.
O
2
calibration in process. Calibrating O2 Analyzer
Message clears when calibration is
complete.
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...