Additional Information
12-8
Equipment Care
Clean and Check the System After Each Use
WARNING:
Turn the system power off before inspection, cleaning, preparation for
use, or storage.
CAUTION:
Do not disinfect with Virex
®
Tb*. Do not use chemical solvents such
as ether, acetone, etc. in or on any part of the system as such solvents may be
destructive to the device. Do not spill anesthetics on the system or components
as chemicals such as Forane (isofluorane) may be destructive.
CAUTION:
Do not immerse the hand cranks.
Follow your institution’s infection control policies and procedures when cleaning
this equipment. Contact your institution’s Infection Control or Safety Coordinator if
assistance is needed. Heed all manufacturer’s warnings concerning the handling of
disinfectants. Work in a well ventilated area when using disinfectants. Disinfectants
should be applied after cleaning with a cloth dampened with warm, soapy water as
organic materials may inactivate chemical germicides and protect microorganisms
from the disinfecting process. Disinfecting solutions should be wiped dry after
approximately 10 minutes of contact time.
After Each Use. . .
Clean and Check your Components
Check the water trap.
Check the water trap, located on the air or 95/5 inlet of the gas
blender, and drain, if necessary.
Clean air, level, and
Clean with mild soap and water solution. Do not use any
flow sensors.
sharp instruments or abrasive materials.
Do not drop the sensors as shock may damage them.
Wipe transducer of air sensor. Do not immerse air sensor.
Between cases store the air sensor on the bracket.
Do not sterilize the air, level, or flow sensors as this will
damage the sensor.
Wipe the level sensor transducer lens to remove gel.
If a sensor is damaged, contact Terumo Customer Service.
Clean the CCM.
Use a cloth dampened with glass cleaner to clean the Central
Control Monitor display screen.
Clean all system surfaces,
Working in a well ventilated area, remove spilled liquids,
including pump heads,
blood, etc.
occlusion rollers, tube clamp
assemblies, and centrifugal
Use a disinfecting solution recommended by your institution.
drive motor.
Wipe all surfaces - except the air, level, and flow sensors. See
sensor cleaning instructions above. Check tube clamp assembly
for spills. If contaminated, follow the cleaning instruction in
Removal and Cleaning of Tube Clamps in this section.
If necessary, use appropriate cleaning pads and polish to
remove residue from stainless steel and aluminum surfaces.
* Virex
®
Tb is a registered trademark of JohnsonDiversy™
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...