Central Control Monitor
4-22
To Change Pump Response:
The bottom portion of the screen displays the configured safety connections and
gives the user the ability to change them for the current case.
1. Select the event/pump response to be changed. The Response button at the
bottom of the screen displays the current pump response.
2. Touch the
Response
button to toggle through the list of available response
choices. The response that is displayed on the button is the response of the
pump for the specified event.
Changes made to pump responses from the Perfusion Screen are only retained for
the current case. The response will revert to the response set in Configuration when
the case is exited.
Occluders Subtab
From this subtab, the user can change an occluder response to an external event
generated by the primary pump. The name of the occluder will be displayed on the
Occluder button.
Occluder responses to the following centrifugal pump events must be the same:
pump stop, pump Coast, and minimum flow. The system will automatically assign the
response chosen for stop, Coast, or minimum flow to the other two events.
To Change Occluder Response:
1. Select the event/response to be changed. The Response button at the bottom of
the screen displays the current occluder response.
2. Touch the
Response
button to toggle through the list of available response
choices. Occluder response choices include No Response, Full Open, Full
Close, or Go to % Flow. The response that is displayed on the button is the
response of the occluder for the specified event. If Go to % Flow is a response
that is selected, touch the
0%
button. Use the arrows to adjust the value. Large
arrows adjust the value by 10% and small arrows adjust the value by 1%.
Changes made to a safety response from the Perfusion Screen are only retained for
the current case. The response will revert to the response set in Configuration when
the case is exited.
Response Button
Setup
Gas
Occluders
Patient
Time/Date
Timers
Pumps
Arterial Pump Stop
Events
Occluder:
Occluder
Full Close
3/8 in.
Arterial Pump Start
Full Open
Arterial Pump Stop
Full Close
Arterial Pump Pause
Full Close
Occluder Response:
16:07:51
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...