System Overview
2-10
ARTERIAL
2010 RPM
L/min
4 . 5 0
Motor speed, flow rate, as well as safety, status, and error messages are displayed
on the centrifugal control unit control panel as well as on the Central Control Monitor.
Pump speed is displayed in RPMs on the pump as well as represented as an amber
colored bar graph.
There are 3 operating modes for a centrifugal pump:
•
Continuous: pump operates at a constant speed set by the user.
•
Pulse: available on the primary pump only; the pump speed alternates between a
peak speed and a baseline speed. The baseline speed can be adjusted through
the Central Control Monitor.
•
Servo Control: pump speed is regulated to maintain a specific pressure or flow
rate. The pressure and flow rate setpoints can be adjusted through the Central
Control Monitor or from the pump speed control knob on the centrifugal control
unit.
The centrifugal system can be connected to safety and monitoring systems and other
pumps during configuration, followed by choosing centrifugal control unit responses.
Centrifugal control unit response choices are No Response, Stop, Coast and
Message Only. The choices allowed for a pump response vary depending on the type
of alert or alarm and the type of device connected to the pump.
When a connected device is in alert or alarm mode, a message will display on the
centrifugal control unit display.
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...