Central Control Monitor
4-74
Place the Flow Sensor
Place the flow sensor by touching a white square. In order for flow information
to be displayed on a centrifugal control unit, a flow sensor must be placed on the
centrifugal pump speed/flow display area. Change the placement of the flow sensor
by touching any other white square.
WARNING:
Always place a flow sensor on a centrifugal pump speed/flow
display area to enable the minimum flow alarm and any safety connections that
have been set for a minimum flow alarm.
Choose
Next
to continue.
Set Safety Connections
After a flow sensor is placed, the system automatically prompts the user to set a
safety connection.
Choose
Yes
to set a safety connection.
Note:
If a centrifugal pump is placed in a position other than in the number 1 or
number 6 position on the CCM, the Speed/Flow Display area will be hidden from view
when in the Edit Safety Connections menu. To avoid moving the pump or flow sensor
in order to make safety connections, make the safety connections at this time.
Choose
No
to defer setting a safety connection.
Do you want to setup a safety connection for ART Flow?
Event
Response
Yes
No
?
L/min
Clamp
Reset
Reset
000:00
Pump
Reset
Reset
000:00
Arterial
Vent
Sucker1
Sucker2
CPG
mmHg
Gas
% O2
FiCO2
0
0
0
To Place the Device 'ART Flow',
Touch One of the Valid Location(s) Above:
Centrifugal Flow Must Be Placed On Pump
Do Not Place
Next >
Next >
< Previous
Undo Place
Undo Place
Speed/flow display
Placing a device icon in this
location may interfere
with a CDI display icon.
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...