Modules
7-4
Air Bubble Detection (ABD)
Enabling, Resetting, or Disabling Air Bubble Detection Alarms
Select the
Air
subtab within the Safety tab on the Perfusion Screen or touch the
Air
icon on the CCM.
1. Touch the air detection
On
button to enable the alarm capability for the air
detection system and the icon on the CCM turns green.
2. Touch the air detection
Off
button to disable the alarm capability for the air
detection system and the icon on the CCM turns gray.
3. When air is detected, correct the condition, and then touch the
Reset
button
on the CCM to clear an air alarm and leave the air bubble detector enabled. If
an air detector has a safety connection to a pump, the alarm can also be reset
by touching the
Select
button on the pump control panel after the air has been
cleared. The ABD is the only safety device that requires a reset by the user to
confirm that the issue has been addressed. All other safety devices automatically
reset when the issue is resolved.
System Response to Air Bubble Detection Alarm
Air
Air
Connected Audible
Icon
Module Message
Device(s) Tone
Flashes red
LED flashes
Message with sensor name
Respond as
Dual-tone
red
in the CCM message area
configured
alarm
and on the local display of any
pumps with an air safety
connection. Safety tab opens.
The alarm state will clear when the condition is corrected and the air bubble detection
system is Reset or turned Off.
Note:
The ABD system will alarm when the air sensor is disconnected from the
module.
Safety
ART Air
Next
Reset
Level
16:07:51
Air
Pressures
Off
On
On Button
Off Button
Reset Button
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...