Central Control Monitor
4-51
Transfer Perfusion Screen
A PC card can be used to transfer Perfusion Screens from one System 1 to another.
Insert a PC card into the System Card slot of the CCM. Use the Move or Copy
function of the PC Card Utilities to copy the desired screen or screens onto the PC
card. Remove the PC card from the CCM and insert into the system slot on the CCM
of another System 1. Use the Move or Copy function of the PC Card Utilities to copy
the desired screen or screens onto the system computer of the other System 1.
If the pumps and devices in a transferred Perfusion Screen were assigned to the
corresponding physical components of the preceding System 1, they must first be
unassigned from those devices and then assigned to the corresponding physical
components on the System 1 the Perfusion Screen was transferred to.
To unassign pumps and devices, access the Configuration functional area, select the
name of the transferred Perfusion Screen, choose “Edit”, choose “Assign Devices
to Modules”. The device types appear in the list under Devices - select a device
type, choose Assign devices, in the top portion of the Assign Configuration Device
to Physical Module window the Assigned Device field will be blank, the device type
appears under the Assigned section in the lower right hand portion of the window–
select the device type from this window and select Unassign – the device type
moves from Assigned to Unassigned. Select OK to return to the Configure Perfusion
Screen-Assign Devices window. Choose each device type in the Devices list in turn
and unassign the remaining pumps and devices. Once all pumps and devices have
been unassigned, follow the instructions in this chapter,
Assign Devices to Modules
section, to assign the pumps and devices to the physical components of the current
System 1. Any safety connections set up for the pumps and devices in a configured
Perfusion Screen, will transfer with that screen when copied to another System 1. Be
sure to check all the safety connections and modify if necessary.
To more easily transfer Perfusion Screens, do not assign the pumps and devices
to the physical components of a particular System 1 when creating the Perfusion
Screen. Transfer the Perfusion Screen to the desired systems and assign the pumps
and devices of each system to the Perfusion Screen.
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...