Centrifugal Control Unit
6-18
Priority Condition
CCU Display Message
Operator Response
Error
Pump stops because:
STOPPED: OVERSPEED
Clamp the lines. Try to restart the
• Motor speed is greater
pump. If the message reappears,
than the maximum
discontinue use of the pump until it has
pump speed OR
been serviced by Terumo
• Motor speed greater
Cardiovascular Systems. Reassign or
for more than 2 seconds.
use equipment to maintain perfusion.
Pump board temperature SERVICE PUMP
Discontinue use of the pump until it
> 80 ºC for more than
has been serviced by Terumo
2 seconds.
Cardiovascular Systems. Reassign or
use standby equipment.
Low level motor fault
STOPPED: MOTOR ERROR Clamp the lines. Try to restart the
pump. If the message reappears,
discontinue use of the pump until it
has been serviced by Terumo
Cardiovascular Systems. Reassign or
use standby equipment to maintain
perfusion.
Overcurrent condition
STOPPED: OVERCURRENT Clamp the lines. Try to restart the
detected.
pump. If the message reappears,
discontinue use of the pump until it has
been serviced by Terumo
Cardiovascular Systems. Reassign or
use standby equipment to maintain
perfusion.
Motor draws too much STOPPED: PUMP JAM
Clamp the lines. Try to restart the
current
pump. If the message reappears,
discontinue use of the pump until it
has been serviced by Terumo
Cardiovascular Systems. Reassign or
use standby equipment to maintain
perfusion.
Self test failure.
SERVICE PUMP
Discontinue use of the pump until it
has been serviced by Terumo
Cardiovascular Systems.
Status Drive Motor has
CONNECT MOTOR
Message should clear when the motor
become disconnected.
is reconnected.
Manual button is
MANUAL: MONITOR FLOW Message should clear when speed is
pressed when the pump
changed or pump is stopped.
is in Pulse mode.
Manual button is pressed MANUAL: MONITOR FLOW Message should clear when speed is
when the pump is in
changed or pump is stopped.
Servo mode.
Stop button pressed in TURN SPEED TO ZERO
Message clears in 4 seconds if no
Continuous mode with
action is taken. Press Stop button
speed > 1550 RPM.
again to stop the pump.
Stop button pressed in TURN SPEED TO ZERO
Message clears in 4 seconds if no
Pulse mode with average
action is taken. Press Stop button
RPM > 1550.
again to stop the pump.
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...