Centrifugal Control Unit
6-4
System Connection
Route and arrange the pump cable to minimize exposure to spills. Connect the
cable to one of the pump connections on either side of the system base or, if
necessary, to any of the module connections. With a clear view of the desired pump
or module connection, carefully connect the pump cable to one of the system pump
connections on either side of the system base or, if necessary, to any of the system
module connections. Do not use excessive force to seat the connector and make
sure not to bend the connector pins. Tighten screws to secure the connection.
Standoffs are provided in the accessories kit. Two standoffs should be installed on a
module receptacle to securely fasten a pump cable or flowmeter cable to the module
receptacle. Only a pump power cable or flowmeter power cable can be inserted into
a receptacle which has the standoffs installed. Modules cannot be inserted into a
module receptacle when standoffs are installed.
Manual Drive Unit
Terumo Cardiovascular Systems Corporation recommends
that either the manual drive unit or a roller pump be prepared
for use as standby equipment. Ensure that the tubing length
is sufficient.
If using the manual drive unit, mount the manual drive unit
securely to a vertical pole, tightening both bracket knobs.
Be sure there is sufficient space around the handle to turn it
easily.
If using a roller pump, position it on either the inlet or the
outlet side of the centrifugal pump. Then adjust the pump
so that the tubing is totally unoccluded.
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...