Roller Pumps
5-31
Priority
Condition
Pump Display Message
Operator Response
Status
Follower pump has
STOPPED: MASTER
Message should clear when
stopped in response to
Master pump is restarted.
master pump event.
Follower pump has
PAUSED: MASTER
Message should clear when
paused in response to
master pump resumes.
master pump event.
Master pump has
STOPPED: FOLLOWER
Message should clear when
stopped in response to
master pump is restarted.
follower pump event.
Master Pump has
PAUSED: FOLLOWER
Message should clear when
paused in response to
follower pump resumes.
follower pump event.
Pump pauses because PAUSED: COVER OPEN
Message should clear when cover
cover has been opened.
is closed.
Pump direction has
PUMPING: REVERSE
Message should clear when default
been changed to
direction is resumed. Default
reverse.
direction is resumed when the
pump stops.
Pulse mode has been
RAMP BASELINE: XX%
Monitor the speed change. The
activated - base is
base is reduced by 5% every 2
ramping down to target
seconds. Message clears when
value.
the target Base setting is reached.
Manual button is
MANUAL: MONITOR FLOW
Message should clear when speed
pressed when the
is changed or pump is stopped.
pump is in Pulse mode
or Servo mode.
Stop button is pressed MANUAL: MONITOR FLOW
Message clears when pump is
in Pulse mode or
started.
Servo mode.
Pump is paused during PAUSED: <current volume> mL Message clears when pump
volume auto dose
resumes or stops.
delivery.
Pump is paused during PAUSED:
Message clears when pump
time auto dose delivery. <current time remaining>
resumes or stops; time continues
to count down.
Volume auto dose
DONE: <current volume> mL
Message clears when auto dose
delivery finished.
delivery is restarted.
Time auto dose
DONE: 00:00
Message clears when auto dose
delivery finished.
delivery is restarted.
Pump stops during
STOPPED: <current volume> mL Message clears when pump is
volume auto dose
restarted; volume is reset to zero.
delivery.
Pump stops during
STOPPED:
Message clears when pump is
time auto dose
<current time remaining>
restarted; time resumes counting
delivery.
down.
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...