System Overview
2-11
Modules
Features
The following modules connect to the system: Air Bubble Detect, Level Detect,
Occluder, Temperature, Pressure, Flowmeter, Interface Module for CDI 100/101,
Interface Module for CDI 500, and Interface Module RS-232.
Modules provide an interface between an appropriate type of sensor or device and
the system.
Each module has appropriate connections for the type of sensor needed.
Each module has a tricolor LED (light emitting diode) to indicate module status.
Status modes are:
Constant Green – Online Mode
Blinking Yellow – Alert Condition
Blinking Green – Assignment Mode
Constant Red – Broken Mode
Constant Yellow – Offline Mode
Blinking Red – Alarm Condition
Up to 18 modules can be connected to the system; 9 on each side.
Modules are connected directly to the receptacles on either side of the system, with
the exception of the flowmeter module, which is mounted on a bracket and connected
to the system by a cable from the module.
System Connection
LED
Sensor Connection