Central Control Monitor
4-58
Configure Perfusion Screen
This screen shows a graphical depiction of the patient and primary pump orientation,
as well as the primary pump type and reservoir type selected.
Two timers, labeled Pump and Clamp, the oxygenator, and the icon for the electronic
gas blender, if installed, appear on the screen. The locations of these devices and
controls are fixed and cannot be changed. The primary pump displays the type,
color, and flow units selected.
The Configure Perfusion Screen is used to add other pumps and devices to the
perfusion circuit. Select an option from the menu.
Configure Other Pumps
Select Configure Pumps to add other pumps and designate a cardioplegia pump and
a backup pump. A total of six pumps can be displayed and controlled on a Perfusion
Screen.
A list of pumps that have been added appears below the menu. The primary pump
name should appear in the list.
Use the
Add
button to add additional pumps.
The Add button is available until the maximum number of pumps have been added to
the Perfusion Screen. Use the
Edit, Place,
and
Remove
buttons to make changes to
pumps that have been added to the Perfusion Screen. Touch the pump name on the
list to activate the appropriate buttons.
Clamp
Reset
Reset
000:00
Pump
Reset
Reset
000:00
Arterial
L/min
START
CABG
Close
Pumps:
Arterial
Assign Devices to Modules
Configure Pumps
Configure Other Devices
Edit Safety Connections
Save As >
Add >
Configure Perfusion Screen > Configure Pumps
Menu
Edit >
Place >
Edit >
Place >
Remove >
Remove >
Gas
% O2
FiCO2
0
0
CABG ROLLER
0
Menu
Pumps Added List
Timers
Add, Edit, Place
and
Remove
Buttons
Primary Pump
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...