Centrifugal Control Unit
6-17
Priority Condition
CCU Display Message
Operator Response
Alert
Pump has reached
AUTO: MAX RANGE
Check servo source. Check perfusion
maximum speed trying
circuit. Alert will clear when speed
to attain setpoint.
slows to within operating range.
Pump has reached
AUTO: MIN RANGE
Check servo source. Check perfusion
minimum speed trying
circuit. Alert will clear when speed
to attain setpoint.
increases to operating range.
Error
CCM missing.
NO SYSTEM COMPUTER
Continue use of centrifugal control unit
from front panel controls and displays.
Be aware that the pump may not
respond to events from connected
devices. Do not use the system again
until it is serviced by Terumo
Cardiovascular Systems Service.
Network
NO COMMUNICATION
Continue use of centrifugal control unit
communication failure.
from front panel controls and displays.
Be aware that the pump may not
respond to events from connected
devices. Do not use the system again
until it is serviced by Terumo
Cardiovascular Systems Service.
Loss of communication CHECK LINKED PUMP
Check for proper operation of the
from connected roller or
connected pump. If message does not
centrifugal pump.
clear, be aware that the pump may not
respond to messages or operations
requiring information from the
connected pump.
Loss of connected air
CHECK AIR SENSOR
Check for proper operation and
bubble detector module.
connection of the module or try
connecting module to another base
receptacle. If message does not clear,
be aware that the pump may not
respond to messages from the module.
Replace module and reassign.
Loss of flowmeter
CHECK FLOWMETER
Check for proper operation and
module.
connection of the module or try
connecting module to another base
receptacle. If message does not clear,
be aware that the pump may not
respond to messages from the module.
Replace module and reassign.
Loss of connected level CHECK LEVEL
Check for proper operation and
module.
connection of the module or try
connecting module to another base
receptacle. If message does not clear,
be aware that the pump may not
respond to messages from the module.
Replace module and reassign.
Loss of connected
CHECK PRESSURE
Check for proper operation and
pressure module.
connection of the module or try
connecting module to another base
receptacle. If message does not clear,
be aware that the pump may not
respond to messages from the module.
Replace module and reassign.
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...