Roller Pumps
5-28
Priority
Condition
Pump Display Message
Operator Response
Alert
Connected centrifugal
< Response >: PUMP COAST
Restart pump if stopped.
pump has gone to
Coast speed.
Connected flow meter
< Response >: BACKFLOW
Clamp the lines.
backflow detected.
Correct backflow condition. Check
for air; clear if present. Restart
pump if stopped.
Alert over pressure
< Response >: OVER PRESSURE
Correct over pressure condition.
detected for connected
pressure sensor.
Low Level detected for
< Response >: LOW LEVEL
Correct low level condition.
connected alert level
sensor.
Follower Pump
CANNOT MAINTAIN RATIO
Reduce the follower ratio setting
unable to maintain
or adjust the master pump speed.
Master ratio setting.
Stop then start master pump. If
message appears again,
discontinue M/F operation.
Follower stops due to
STOPPED: NO MASTER
Check master pump cable
loss of data from
connections then stop and start
Master pump.
master pump. If message appears
again, reassign the master pump.
Do not use the master pump until
it has been serviced by Terumo
Cardiovascular Systems.
Master stops due to
STOPPED: NO FOLLOWER
Check the follower pump cable
loss of data from
connections. Stop and start the
Follower pump.
master pump. If message appears
again, reassign the follower pump. Do
not use the follower pump until it has
been serviced by Terumo Cardiovascular
Systems.
Loss of servo source
MANUAL: MONITOR FLOW
Check the operation and placement
data.
of servo source. If source is
operating properly, try to initiate
servo operation via the CCM. If
message appears again, discontinue
use of servo operation.
Cannot maintain servo
MANUAL: MONITOR FLOW
Check the operation and placement
setpoint.
of servo source. If source is
operating properly, try to initiate
servo operation via the CCM. If
message appears again, discontinue
use of servo operation.
Motor speed < demand
UNDERSPEED
speed for more than 2
seconds.
At start-up, if you are operating under
cold temperatures and heavy occlusion
(see Resolving Pump Jam Section) an
underspeed warning may occur. The
underspeed message will clear as the
pump rotates and the tubing is warmed.
If the message does not clear within 5
minutes, stop and start the pump. If the
message reappears, contact Terumo
Customer Service. Reassign, use
standby equipment, or use hand crank.
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...