Modules
7-26
Flow Monitoring
4. Close the sensor latch. Ensure the latch closes securely.
The flow system is operational when Perfusion Screen has initialized and the flow
icon displays “0.00” L/min.
Once a flow sensor is connected to the circuit, it does not need to be calibrated or
turned on to enable flow monitoring. Once there is blood or prime solution circulating
in the system, verify the flowrate display by varying the speed and seeing that the
display changes accordingly.
System Response to Alarm Condition
The flow system will alarm in a backflow condition.
Flow
Flow
Connected Audible
Icon
Module
Message
Device(s)
Tone
Flashes red
LED flashes
Message with sensor name
Respond as
Dual-tone
red
in the CCM message area and
configured
alarm
on the local display of any pumps
with a flow safety connection.
L/min
0.00
Safety
ART Flow
Previous
Flow
0.00 L/min
Temps
16:07:51
Safety
ART Flow
Previous
Flow
Temps
16:07:51
0.62 L/min
ART Flow : BACKFLOW