Technical Support
11-15
Nominal AC (Mains) Power Consumption
System
Configuration*
100 VAC
115 VAC
120 VAC
220 VAC
240 VAC
4-Pump System 9.5A – 11.5A
8.5A – 10A
8A – 10A
4.5A – 5A
4A – 5A
5-Pump System 10.5A – 13A
9.5A – 11.5A
9A – 11A
5A – 6A
4.5A – 5.5A
6-Pump System 11.5A – 14A
10.5A – 12.5A
10A – 12A
5.5A – 6.5A
5A – 6A
7-Pump System 13A – 15.5A
11.5A – 14A
11A – 13.5A
6A – 7A
5.5A – 6.5A
8-Pump System
14A – 17A
12A – 15A
12A – 15A
6.5A – 8A
6A – 7A
*
Each System Configuration is defined as:
All large roller pumps, speed set to 250 RPM while using 1/2 inch (12,7mm) I.D. x 3/32 inch (2,4mm)
wall tubing, properly occluded and fluid temperature of 25°C;
All available module connections used with any combination of Flow, Temperature, Pressure, Level, Occluder,
Air Bubble Detector, and Serial Interface Modules with all assigned and activated;
1 Electronic Gas Blender Module, connected and activated;
1 Lamp connected and powered “ON”;
1 Central Control Monitor, connected and activated;
Peripheral equipment connected to the auxiliary outlet and activated.
This device requires a dedicated 20 Amp mains power source for the 100V - 120V
model or a dedicated 10 Amp AC (Mains) power source for the 220V - 240V model.
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...