Central Control Monitor
4-73
Add Flow Sensors
A maximum of 4 flow sensors is allowed in a Perfusion Screen.
1. Select
Add
.
2. Select Flow Sensor as device type.
3. Choose name from the list or choose
Enter New Name
and use the keyboard to
enter a device name.
Do you want to setup a safety connection for ART Pres?
Event
Response
Yes
No
?
mmHg
Primary Pump requires Pressure Sensor Alert and Alarm Connections.
Choose
Undo Place
to remove the pressure sensor from the Perfusion Screen.
Choose
Next
to continue.
Set Safety Connections
After a pressure sensor is placed, the system automatically prompts the user to set
a safety connection. Alert and alarm safety connections are required between a
pressure sensor and the primary pump.
Choose
Yes
to set a safety connection.
Choose
No
to defer setting a safety connection.
Or Enter New Name:
Enter New Name >
To Add Device:
Cancel
Current defined:
0
Maximum allowed
4
Air Detector
Level Detector
Select Device Type:
Occluder
Pressure Sensor
Flow Sensor
Temperature Sensor
CDI 100
CDI 500
Next >
VEN Flow
CPG Flow
FLOW1
FLOW2
Select From List:
Select Device Name:
Art Flow
Art Flow
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...