Central Control Monitor
4-68
Add a Level Detector
Alert Only and Alert/Alarm
A maximum of 1 level detector can appear on a Perfusion Screen, and it can be
configured for either Alert/Alarm requiring two sensors or Alert Only requiring one
sensor.
To add a Level Detector
1. Select
Add
.
2. Select Level Detector as device type.
3. Name the Level Detector.
4. Select an operating mode for the Level Detector. The default selection is Alert/
Alarm mode.
• Alert Only
allows the user to set up one level sensor on a reservoir. Alert
sensors cannot be used to stop pumps.
• Alert/Alarm
allows the user to set up two level sensors on a reservoir. Alarm
sensors can be used to stop any type of pump or Coast the centrifugal pump.
Touch
Next
to continue.
Or Enter New Name:
Enter New Name >
To Add Device:
Cancel
Current defined:
0
Maximum allowed
1
Air Detector
Level Detector
Select Device Type:
Occluder
Pressure Sensor
Flow Sensor
Select From List:
Temperature Sensor
CDI 100
CDI 500
Next >
Select Device Name:
Level
VENLevel
CPGLevel
LEVEL1
LEVEL2
Level Detection:
Alert/Alarm
Alert Only
Alert/Alarm
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...