Centrifugal Control Unit
6-19
Priority Condition
CCU Display Message
Operator Response
Status Stop button is pressed TURN SPEED TO ZERO
Message clears in 4
in Servo mode with
seconds if no action is
average RPM > 1550.
taken. Press Stop button
again to stop the pump.
Pulse mode has been
RAMP BASELINE: XX%
Monitor the speed change.
activated – base is
The base is reduced by 5%
ramping down to
every 2 seconds. Message
target value.
clears when the target Base
setting is reached.