Central Control Monitor
4-3
Start-Up Messages
If the Power On Self Test fails, the results are displayed continuously at the top of the
Main Screen as long as the Main Screen is displayed or the condition exists. The
failure message is displayed in the following format:
<Module Type Name>: X “Fail”,
where X is the number of modules of this type that failed their Power-On Self-Test.
A maximum of 8 failure messages can be displayed simultaneously. Possible
messages are as follows:
“System Power”
“Temperature Sensors”
“System Computer”
“Gas System”
“System Network”
“CDI 100”
“Air Detectors”
“CDI 500”
“Level Detectors”
“Large Roller Pumps”
“Occluders”
“Small Roller Pumps”
“Pressure Sensors”
“Centrifugal Pumps”
“Flow Sensors”
“Data Transfer”
The roller pump default direction is set to the system direction during start-up. For
more information on setting system direction, see
Configuration, System Settings
in this chapter.
In the event of a System Power, System Computer, System Network, or Gas System
start-up failure, cycle power by turning the Mains Power off and then back on. If the
message is still displayed, contact Terumo Customer Service.
For a roller pump, centrifugal pump, or flow module start-up failure, disconnect and
then reconnect the cable from the back of the pump or module. If the message
appears again, follow the module reassign procedure in this chapter,
System tab
,
Reassign Module
subtab
, and contact Terumo Customer Service.
For all other start-up failure messages, disconnect and then reconnect the module
to the module receptacle. If the message appears again, follow the module reassign
procedure in this chapter,
System tab
,
Reassign Module subtab
, and contact
Terumo Customer Service.
In the event of a CCM start-up failure, the CCM may lose communication capability
with the system. The message “No System Computer” will be displayed on the
control panels of the roller pumps and centrifugal control unit and the message
“System Computer Needs Service” may appear on the CCM.
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...