Central Control Monitor
4-21
To Change Pump Response:
The bottom portion of the screen displays the configured safety connections and
gives the user the ability to edit them for the current case only.
1. Select the event/pump response to be changed. The Response button at the
bottom of the screen displays the current pump response.
2. Touch the
Response
button to toggle through the list of available response
choices. The response that is displayed on the button is the response of the
pump for the specified event.
Changes made to pump response from the Perfusion Screen are only retained for the
current case. The response will revert to the response set in Configuration when the
case is exited.
For Centrifugal Pumps
Touch the
Pump
button until the correct name is displayed on the button. Changes
can be made while the pump is operating.
Change Display Units:
Touch the
Display Units
button to toggle through the choices. Stop when the desired
selection is displayed. Changes are retained for the current case only.
Set Min Flow Alarm:
The Min Flow Alarm button is only available when a flowmeter has been placed on
the centrifugal pump icon. The range for the Min flow alarm is 0.0 to 8.0 L/min.
1. Touch the
Min Flow Alarm
button.
2. Use the arrows to adjust the minimum flow value. The large arrows adjust the
value by 1 L/min and the small arrows adjust the value by 0.1 L/min.
3. Touch the
Min Flow Alarm
button again to store the new value.
Setup
Gas
Display Units:
Occluders
Patient
Time/Date
Timers
Pumps
Pump Response:
Events
Pump:
Arterial
Type:
L/min
Centrifugal Pump
Min Flow Alarm
(L/min):
Level Level Alert
Message Only
Level Level Alarm
Stop
ART Air Air Alarm
Stop
Level Level Alarm
Stop
1.0
16:07:51
Only 3 events can be viewed at
one time. Use scroll arrows to
view other configured events
and pump response.
Response Button
Pump Button
Device Name
and Event
Response
Display Units Button
•
RPM
•
L/min
•
L/min/m
2
•
mL/kg/min
Primary Pump
only with a
configured
flowmeter
{
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...