Central Control Monitor
4-38
Central Control Monitor Messages
The following table includes all the messages that can appear on the CCM display
and corresponding operator responses.
Priority
Condition
CCM Display Message
Operator Response
Alarm
Battery is depleted
SYSTEM POWER
Use manual backup equipment to
and system is shutting
SHUTDOWN
finish case. Fully recharge battery
down.
before using the system again.
Pump goes to 0 flow as
< Pump Name >: STOPPED Correct condition.
a result of a Reduce
Start pump if stopped.
Speed response to an
alert or alarm.
Pump goes to 0 flow as
< Pump Name >: STOPPED Correct the conditions that caused
a result of receiving a
the Reduce Speed/Pause responses.
Pause and a Reduce
Restart the pump.
Speed response.
Air detected.
< ABD Name > :
Correct air condition, reset air sensor,
AIR DETECTED
restart pump if stopped.
Air detected.
AIR DETECTED:
Correct air condition, reset air sensor,
< Pump Name >
restart pump if stopped.
< Pump Response >
Air sensor on, and
< ABD Name > :
Reconnect air sensor, reset air
disconnected.
AIR DETECTED
sensor, restart pump if stopped.
Pressure has risen
< Pressure Name > :
Correct over pressure condition.
above alarm limit.
OVER PRESSURE
Start pump if stopped.
Pressure has risen
OVER PRESSURE:
Correct over pressure condition.
above alarm limit.
< Pump Name >
Start pump if stopped.
< Pump Response >
Reservoir level below
< Level Detector Name > : Fill reservoir to above alarm level.
alarm level.
LOW LEVEL
Alarm clears automatically.
Restart pump if stopped.
Reservoir level below
LOW LEVEL:
Fill reservoir to above alarm level.
alarm level.
< Pump Name >
Alarm clears automatically.
< Pump Response >
Restart pump if stopped.
Backflow has been
< Flow Sensor Name > : Correct backflow condition.
detected.
BACKFLOW
Alarm clears automatically.
Restart pump if stopped.
Backflow has been
BACKFLOW:
Correct backflow condition.
detected.
< Pump Name >
Alarm clears automatically.
< Pump Response >
Restart pump if stopped.
Flow for flow sensor
< Pump Name > :
Correct low flow condition.
placed on a centrifugal
MINIMUM FLOW
Alarm clears automatically.
pump has gone below
minimum limit.
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...