System Base
3-18
Set Screws
To Install or Remove Extensions:
1. Remove cable cover.
2. Remove pumps.
3. Remove module cover.
4. Loosen set screws on each side of the base. Some modules may need to be
removed to access the set screws.
5. Slide endcaps out.
6. Insert or remove the extension.
•
If removing extensions, remove securing screws and remove extension.
•
If adding extension, insert extension and install securing screws.
7. Push the end caps in until they stop.
8. Tighten set screws. Set screws must be securely tightened to ensure pump
stability during operation.
9. Install pumps according to the instructions in
Roller Pump
chapter.
10. Install module covers and cable cover according to the instructions at the end of
this chapter.
Securing Screws