Central Control Monitor
4-20
To Calibrate O
2
Analyzer:
Touch the
Calibrate
button.
•
Calibration takes approximately 2 minutes.
•
“Calibrating O
2
analyzer” is displayed during calibration.
When the calibration is complete, the message “O
2
Analyzer Calibration Complete”
will display in the Messages subtab within the Aux tab and in the message area of the
CCM if there are no higher priority messages.
Pumps Subtab
The Pumps subtab allows the user to change the roller pump display units, tube size,
and pump response to a safety event. The Pumps subtab also allows the user to
change centrifugal pump display units, pump response to a safety event, and set the
minimum flow alarm. See the
Roller Pumps
or
Centrifugal Control Unit
chapter for
detailed information about connecting and operating pumps.
For Roller Pumps
Touch the
Pump
button until the correct name is displayed on the button. The roller
pump must be in Stop mode to change the display units and the tube size.
To Change Display Units:
Touch the
Display Units
button to toggle through the selections. Stop when the
desired selection is displayed. Changes are only retained for the current case.
To Change Tube Size:
1. Use the scroll arrows to view the list of available sizes.
2. Touch the correct tube size from the list. The tube size selection is highlighted
and the tube size will display above the selection box and on the roller pump local
display.
To enter a custom tube size:
1. Select Custom from the list.
2. Touch the
Flow Constant
button.
3. Use arrows to adjust value.
4. Touch button again to retain the value.
The system will retain this selection until a new selection is made from this
screen or from the pump local display.
Display Units Button
•
L/min
•
RPM
•
L/min/m
2
•
mL/kg/min
Setup
Gas
Display Units:
Occluders
Patient
Time/Date
Timers
Pumps
Pump Response:
Events
Pump:
Arterial
Type:
Large Roller Pump
Tube Size:
Custom
Custom
S4:1
S8:1
ART Pres Pres Alert
Reduce Speed : 300 mmHg
ART Pres Pres Alarm
Stop : 360 mmHg
ART Flow Back Flow
Stop
ART Pres Pres Alarm
Stop
16:07:51
FC144
Flow Constant
L/min
L/min
Response Selection Button
Device Name
and Event
Response
Tube Size Box
Pump Button
6mm
8mm
10mm
12mm
S1:1
S2:1
S4:1
S8:1
Custom
Only 3 events can be viewed
at one time. Use scroll arrows
to view other configured events
and pump response.
Primary
Pump Only
}
1/8 in.
3/16 in.
1/4 in.
5/16 in.
3/8 in.
7/16 in.
1/2 in.
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...