Roller Pumps
5-3
Installation and Connection
Pump Cable
Connect the pump cable to the receptacle located on the back of the roller pump.
Align the red dot on the pump cable connector with the red dot on the pump
receptacle. Insert the connector into the receptacle; an audible click indicates that the
connector is fully seated in the receptacle. Pull gently on the cable to verify a secure
connection.
Base Mounting
Ensure the mounting lever is in the up position. Grasp roller pump firmly with two
hands. Tilt pump forward and insert the mounting tab into the front mounting base.
Lower the rear of the pump onto the mounting rail. Ensure the mounting lever clicks
into place and the pump is secure on the rail.
Mounting Rail
Mounting Lever
To System Base
Red Dot
Summary of Contents for 801763