Technical Support
11-10
Pressure
Part
of
Mnemonic Description
Value
Units
Type
Snapshot?
OPAL
Over pressure alert
Safety Event
OPLC
Over pressure alert cleared
Safety Event
OPAM
Over pressure alarm
Safety Event
OPMC
Over pressure alarm cleared
Safety Event
PRES
Average pressure
-250 to 900
mmHg
Periodic
Yes
Pumps
Part
of
Mnemonic Description
Value
Units
Type
Snapshot?
MLPM
Flow
-10000 to 10000 mL/min
Periodic
Yes
CIF
Cardiac index flow
0.0 to 10.0
L/min/m2
Periodic
Yes
KGF
Kilogram flow
0.0 to 10000
mL/min/kg
Periodic
Yes
PDIR
Direction
CW or CCW
Event
POCU
Overcurrent
Safety Event
POCC
Overcurrent cleared
Safety Event
POSP
Overspeed
Safety Event
POSC
Overspeed cleared
Safety Event
PPAU
Pump paused (only for roller pumps)
Safety Event
PRAT
Ratio (when follower)
1:1, 2:1, 4:1,
M:F
8:1, 16:1
Periodic
Yes
0 to 99
%
PSTA
Pump started
Safety Event
PSTO
Pump stopped
Safety Event
PUSP
Underspeed
Safety Event
PUSC
Underspeed cleared
Safety Event
RPM
RPM
0 to 250
RPM
Periodic
Yes
0 to 4000
TMLP
Total flow
0 to 20000
mL/min
Periodic Yes
System
Part
of
Mnemonic Description
Value
Units
Type
Snapshot?
SOFB
Off battery (on AC)
Safety Event No
SONB
On battery
Safety Event No
TROB
Time remaining on battery 0 to 1440
Min
Periodic
Yes
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...