Modules
7-30
Occluder Module
Insert Tubing
1. From the Perfusion Screen, touch the
Occluder
icon on the CCM. This will
activate the occluder control slide bar.
2. On the occluder control slide bar, touch the
Open
button to retract the plunger sufficiently to insert
tubing.
3. Unlatch and open the occluder cover.
4. Insert tube into the tubing channel of the cover.
5. Close the cover, keeping the tubing properly aligned
in the channel, and secure latch.
Note:
If the cover does not close easily, touch the
Open
button again to retract the
plunger another step. The
Open
button can be pressed up to 3 times.
CAL
Open Button
CAL
CAL
OCCL1