Central Control Monitor
4-2
CCM Bracket
CCM Power Cable
Connect CCM
Attach the CCM bracket to a system pole that will provide easy viewing and access to
the CCM while running a perfusion case.
Make sure AC (Mains) power is OFF before connecting or disconnecting the CCM.
Attach the CCM power cable to one of the receptacles located on the left or right side
of the base, being careful not to damage the connector pins.
Before use, check that the CCM is within reach and positioned for maximum visibility.
The display should be positioned within a viewing angle of ± 45° horizontal and ± 30°
vertical; exceeding these viewing angles will decrease the contrast of various colors.
When power to System 1 is turned on, the CCM, the entire system, and attached
modules perform self tests.
CAUTION:
Use care when handling the Central Control Monitor as the
controls, displays, or circuitry may be damaged if the display is dropped,
struck, or immersed in liquid.
CAUTION:
Pump control panels and Central Control Monitor are intended
for hand and gloved hand operation only. Do not use any instruments on
the pump control panels or Central Control Monitor as this may damage the
device.
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...