Central Control Monitor
4-17
Setup
Safety
CPG
Aux
System
Post Case
Perfusion Screen Tabs
The area of the Perfusion Screen below the pump control boxes displays the 6
Perfusion Screen tabs: Setup, Safety, CPG, Aux, System, and Post Case. An
additional tab, the Pump tab, is accessed by touching the pump icon. Pressing any
one of these opens a screen. Each screen may contain a number of subtabs, which
open additional screens that allow access to more information and functions for
individual sensors and devices.
Pump and Clamp Timers
A pump and clamp timer are always available on the Perfusion Screen. A Perfusion
Screen can be configured to include up to 2 additional timers, for a maximum of 4
timers. The additional timers can be configured to count up from zero or to count
down from a chosen value. Count up timers have a range from 00:00 to 990:59;
count down timers have a range from 990:59 to 00:00. The pump and clamp timer
can only be used in count up mode.
To operate a timer:
1. Start the timer by touching the Time Display area. The timer starts counting in
minutes and seconds. The start time appears below the Time Display area.
2. To stop the timer, touch the Time Display area again. The stop time appears to
the right of the start time. The Reset button is activated.
3. To restart the timer, touch the Time Display area again. The timer continues
counting from the current value. The stop time disappears.
4. To reset a count up timer to 0 or a count down timer to its initial value, touch the
Reset
button. The start and stop times disappear.
Clamp
Reset
Reset
000:00
Pump
Reset
Reset
000:00
Time Display Area
•
Estimated battery time remaining in minutes displayed above the Power Source/Battery
Status icon when on battery. An AC (Mains) Power Indicator is shown over the Battery
Status icon when on AC (Mains) Power.
•
Estimated battery time remaining as indicated by icon color:
• When on AC (Mains) Power:
• Steady green indicates a battery system with greater than one hour run time
with a fully loaded system
• Steady red indicates a battery system with less than one hour run time with a
fully loaded system.
• When on Battery Power:
• Flashing green indicates 30 minutes or more battery life at present discharge rate
• Flashing yellow indicates 15 minutes or more, but less than 30 minutes battery life
at present discharge rate
• Flashing red indicates less than 15 minutes battery life at present discharge rate
AC (Mains) Power
Indicator
53 min
WARNING:
If the Battery Capacity is indicated at less than 60 minutes battery
power, do not start a case. If the system should be fully charged, but the Battery
Capacity indicates less than 60 minutes, call Terumo Customer Service. Do not
use the system until it has been serviced.
On Battery Indicator
53 min
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...