Central Control Monitor
4-50
In the center are the utilities buttons. These buttons are used to move, copy, view
summaries, rename, or delete the Perfusion Screens that appear on either list. The
utilities buttons are activated by selecting a Perfusion Screen.
There must be a System PC card installed in the Central Control Monitor for the move
and copy utilities to activate. Refer to the
Central Control Monitor
chapter
PC Card
Compartment
section for instructions on inserting a PC Card.
Move and Copy
Perfusion Screens can be copied and moved between the CCM and PC card or vice
versa. The Copy function is used to back up Perfusion Screens and for placing the
same Perfusion Screen on multiple systems. The Move function removes a Perfusion
Screen from its current location, either system computer or PC card, and moves it to
the other location.
To move a Perfusion Screen
1. Select the Perfusion Screen.
2. Touch the appropriate
Move
button.
Name:
Ready: Last Edit:
Close
PC Card Utilities
Card:
Edit Card Name > >
System Computer
Name:
Ready: Last Edit:
Name:
Ready:
Last Edit:
Sample
Move > >
< < Move
Copy > >
< < Copy
View Summary
Rename
Delete
Move > >
< < Move
Copy > >
< < Copy
View Summary
Rename
Delete
Min AVR
Yes
05/15/2002
Circ-ARREST
Yes
05/15/2002
Pediatric
Yes
05/15/2002
CABG Centrif
Yes
05/15/2002
CABG Roller
Yes
15/10/2002
•
The Accessing PC Card and Building PC Card List windows are displayed in
succession.
•
The chosen Perfusion Screen no longer appears on the original list and appears
on the list on the opposite side.
To copy a Perfusion Screen:
1. Select the Perfusion Screen.
2. Touch the appropriate
Copy
button.
•
The Accessing PC Card List and Building PC Card List windows are temporarily
displayed in succession.
•
The Perfusion Screen is copied and the name appears on the appropriate list.
Perfusion Screen names must be unique. If the user tries to move or copy Perfusion
Screens to locations where a screen of the same name already exists, the user will
be prompted to either overwrite the existing screen or cancel the move or copy action.
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...