Central Control Monitor
4-18
Perfusion Screen Tabs
Setup Tab
The Setup tab contains six subtabs, which allow the user to enter patient data, control
various functions of the gas system, pumps, occluders, and timers, as well as setting
the system time and date.
Patient Subtab
Patient height and weight are entered during the case setup and the patient’s body
surface area and blood volumes are calculated automatically. The patient data
is retained for the current case only and is reset each time a Perfusion Screen is
opened.
To Adjust the Height:
1. Touch the
Height
button. The Height button changes to white.
2. Use the arrows to enter patient height in cm. The large and small arrows initially
adjust the value to 30 cm. Subsequent adjustments are 20 cm for the large
arrows and 1 cm for the small arrows. The maximum height is 250 cm.
3. Touch the
Height
button again to store the new value.
To Adjust the Weight:
1. Touch the
Weight
button. The Height button changes to white.
2. Use the arrows to enter patient weight in kg. The large arrows adjust the value
by 2 kg up to 20 kg, then adjust the value by 20 kg increments. The small arrows
adjust the value by 0.1 kg or 1 kg and correspond to the large arrow increment
value. The maximum weight is 300 kg.
3. Touch the
Weight
button again to store the new value.
Setup
Gas
16:07:51
Occluders
Time/Date
Timers
Pumps
Patient
Subtabs
Setup
Gas
16:07:51
Height
Occluders
Patient
Time/Date
Timers
Pumps
Dubois
m
2
Body
Surface Area
Weight
mL
Blood Volume:
cm
kg
Choices are:
Dubois
Boyd
Pediatric
Arrows
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...