Roller Pumps
5-17
Change Pump Default Direction:
Use one of the methods below to change pump default direction.
•
Plug the pump into the system, power the system on and allow the start-up
sequence to complete. Verify that the pump default direction arrow now matches
the system direction.
•
Pump default direction is also changed to match the system direction during
the reassignment process, except when the pump is running at the initiation of
reassignment. After reassignment, verify that the pump default direction arrow
now matches the system direction.
•
The pump direction can be changed by entering Configuration, assigning the
stopped pump to a Perfusion Screen and then opening that screen from the
Perfusion functional area. Verify that the pump default direction arrow now
matches the system direction.
Pump Menu
The pump menu can be used to temporarily reverse the forward direction, enter a
flow constant, or change the tube size setting.
To access the pump menu:
1. Place the pump in Stop mode. The primary display should be blank.
2. Ensure that there are no active alarms.
3. Press
Select
button to access the main Pump Menu.
4. Turning the pump speed control knob will successively highlight the different
menu options:
• SELECT: DIRECTION
• SELECT: FLOW CONSTANT
• SELECT: TUBE SIZE
• SELECT: EXIT
5. Press the
Select
button to edit the highlighted parameter.
MENU
NO CHANGE
SELECT: EXIT
1/2
Pump Speed
Control Knob
Select
Tube Size or
Flow Constant
Menu Options
Default Forward
Direction
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...