Central Control Monitor
4-26
Deliver Mode
To start tracking cardioplegia delivery:
1. Select Deliver
mode by pressing the
Deliver
button.
2. Start the cardioplegia pump. The cardioplegia pump must be in L/min mode for
total flow rate to be displayed on the CCM and pump local display.
•
The system tracks and displays the volume for Current Delivery and Total Delivery.
•
The Total Delivery value is grayed out indicating that it cannot be modified.
To stop cardioplegia delivery:
1. Touch the cardioplegia pump
Stop
button.
•
Volume tracking stops.
•
Time Since Last Delivery is set to 0 and the timer starts counting the time since
last delivery.
•
If the subtab is closed while the timer is active, the time since last delivery is
displayed on the CPG tab on the CCM.
•
The Total Delivery value is displayed in black indicating that it can be modified.
To modify the Total Delivery volume:
1. Touch the
Total Delivery
button. The button background turns white.
2. Use the arrows to adjust the total volume. The small arrows change the volume
by 10 mL, the large arrows change the volume by 100 mL.
3. Touch the
Total Delivery
button again to store the new Total Delivery value.
•
The adjustment arrows disappear and the Total Delivery button background turns
gray.
Setup
Safety
CPG
Aux
System
Post Case
001:18
CPG
Auto Dose
Setup
Delivery
Time Since Last Delivery:
000:00
mL
Current Delivery:
0
0
mL
Deliver:
Recirculate:
Total Delivery:
16:07:51
CPG
Auto Dose
Setup
Delivery
Time Since Last Delivery:
001:18
mL
Current Delivery:
1000
1000 mL
Deliver:
Recirculate:
Total Delivery:
16:07:51
CPG
Auto Dose
Setup
Delivery
Time Since Last Delivery:
001:18
mL
Current Delivery:
1000
1000 mL
Deliver:
Recirculate:
Total Delivery:
16:07:51
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...