Roller Pumps
5-24
To Initiate Servo Mode:
1. Touch the
Pump
icon to open the Pump tab.
2. Open the
Servo
subtab.
3. Select the source pressure or flow.
4. Touch the
Auto
button.
The pump control panel shows the Servo mode symbol next to the current servo
setpoint value. The pump secondary display shows the current setpoint value.
The pump will respond to all safety connections, including alerts or alarms that have
been established for the servo control source while operating in Servo mode.
To change setpoint:
To change the servo setpoint, turn the pump speed control knob or adjust the pump
control slide bar. Increase/decrease the pump speed. When the knob or slide bar
adjustment stops, the pump locks onto the current average value of the servo source
as the new servo setpoint.
To End Servo Mode:
Use one of the following two options:
•
Press the
Manual
button on the pump front panel and the pump transitions to
Continuous mode at its current speed.
•
Open the
Servo
subtab on the
Pump
tab on the CCM, touch the
Manual
button,
and the pump transitions to Continuous mode at its current speed.
Set Point Value
ARTERIAL
L/min
3/8
4.50
155 mmHg
Servo Symbol
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...