System Base
3-14
The instrument tray in front of the pump platform provides storage space for items up
to a maximum of 10 pounds (4,5 kg).
There are dedicated brackets for many of the components used with the Terumo®
Advanced Perfusion System 1. Refer to specific chapters for instructions on how
to attach pole mount brackets for: roller pumps, centrifugal control units, flowmeter
modules, occluder head, and the Central Control Monitor (CCM).
Loading Zones
To prevent tipping, only transport the system in the lengthwise direction. Do not
transport the system up or down an incline greater than 10 degrees. Do not hang
more than 10 lbs. (4,5 kg) total from the hooks on each telescoping pole. A minimum
of two roller pumps must be placed on the system base pump platform. Distribute the
weight of additional system components and accessories according to the following
figures and instructions.
The System 1 Base includes four poles for mounting equipment and a pole-mounted shelf for
holding loose items. When mounting equipment, it is important not to overload the system or
to concentrate the weight where it could cause the system to tip. The mounting area is divided
into six zones, with limits on the amount of that weight can be mounted in each zone. The
table below shows the limits for each zone.
The drawings below show the location of each zone. Some of the zones overlap (e.g., zones A
and E). If you mount equipment in the overlap area, its weight counts against the limits for both
zones (i.e., it is double counted). To help you calculate the loads, the table after the drawings
shows the weight of various pieces of equipment.
Some points to keep in mind:
•
For stability, place at least two roller pumps on the base
•
For calculating load, the center of the weight of the equipment determines which zone it is
in, regardless of which pole the equipment is mounted on
•
Area D is for the CCM only
•
The load in Area C includes both equipment mounted in the area and the weight of
anything placed on the shelf. The weight of the shelf is not included.
•
The weight of fluids contained in the equipment (e.g., a reservoir) must be included in the
load calculations
Zone
Weight
A
90 lbs. 40,8 kg
B
90 lbs. 40,8 kg
C
50 lbs. 22,7 kg
D
CCM and Mounting Arm
E
90 lbs. 40,8 kg
F
90 lbs. 40,8 kg
Note:
The dimensions noted in the following graphics refer to inches and (meters).
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...