Central Control Monitor
4-98
Edit a Perfusion Screen
To change an existing Perfusion Screen, select the Perfusion Screen from the list and
touch the
Edit
button.
Add, Edit, Place, and Remove
Use the
Add, Edit, Place,
and
Remove
buttons to make changes to the pumps and
devices. Select the component by touching its name on the list. Primary pumps may
only be edited; they may not be removed and their position in the graphic display
cannot be changed. Devices may be added up to a maximum of 24 components,
including 6 pumps in each Perfusion Screen.
Edit
is used to change the cardioplegia or backup designation of a pump, the name
of the pump, and the color of the pump. If the pump type must be changed, remove
the existing pump, and add a new pump of the correct type to the configuration. Edit
is also used to change the name of the device, the color of the device, and properties
of the devices such as pressure and temperature limits.
Place
is used to change the location of the pump or device on the graphic display.
Locations for the primary pump, the CDI devices, and Pump and Clamp timers are
fixed and cannot be changed.
Remove
is used to remove a pump or device from the graphic display. The primary
pump cannot be removed.
New
Edit
View Summary >
Rename >
Set Password >
Delete
Return to Main
Name
Ready:
Last Edit:
Password:
Perfusion Screens
15/10/2002
15/10/2002
CABG Roller
CABG Centrif
Yes
Yes
Min AVR
Yes
Circ-ARREST
Yes
Pediat-
ric
No
Password
15/10/2002
15/10/2002
15/10/2002
Edit Button
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...