Modules
7-34
Interface Module for CDI™ 100/101 and CDI™ 500
3. To connect a CDI™ 100 monitor, connect the serial cable to the serial
communications connector of the monitor.
SAT %
HCT %
8.0 g/dL
HB
75
24
4. Route cable to minimize exposure to spills.
5. Configure the CDI™100 monitor:
•
Turn the monitor on.
•
Enter the setup mode from the main menu.
•
Select RS-232 from the Select Function menu.
•
Set the desired data output interval. Choices are 0, 30, 60 or 120 seconds.
Selecting 0 results in no data being transferred.
•
Select Enter to store a new choice into memory.
•
Select Return to return to the main menu and make no change to the output
interval.
6. To connect a CDI™ 101 monitor, install the gender changer on the serial
communications port of the monitor. Connect serial cable to the gender changer.
7. Route cable to minimize exposure to spills.
8. Configure the CDI™101 monitor:
•
Turn the monitor on.
•
Use setup screen 6 on the monitor to enter these settings:
o Output type:
S100
o Data Bits:
8 data bits, 1 stop bit
o Parity:
None
o Baud Rate:
4,800
o Output Interval:
1 second
Data Transfer
Data is available when a Perfusion Screen is opened on
the CCM.
The icon will display “— — —” when no data is received
or the data is below range.
The icon will display “999” when the data is above range.
On Switch
CDI™ 101
CDI™ 100
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...