Modules
7-13
Level Monitoring
Level Monitoring Messages
Priority Condition
CCM Display Message
Operator Response
Alarm Reservoir level below
< Level detector name > :
Fill reservoir to above alarm level.
alarm level.
LOW LEVEL
Alarm clears automatically.
Restart pump if stopped.
Reservoir level below
LOW LEVEL:
Fill reservoir to above alarm level.
alarm level.
< Pump Name >
Alarm clears automatically.
< Pump Response >
Restart pump if stopped.
Pump goes to 0 flow as < Pump name > : STOPPED
Fill reservoir above alert level.
a result of a Reduce
Restart pump.
Speed response to a
level alert/alarm.
Alert
Reservoir level below
< Level detector name > :
Fill reservoir above alert level.
alert level.
CHECK LEVEL
Start pump if stopped. Alert clears
automatically.
Level sensor alert sensor CHECK LEVEL:
Fill reservoir above alert level.
detects low level.
< Pump Name >
Start pump if stopped. Alert clears
< Pump Respose >
automatically.
Level system is on, alarm < Level detector name > :
Mount the alarm or alert level
level sensor or alert level NOT ATTACHED
sensor to venous reservoir. Alert
sensor not mounted to
message clears automatically.
venous reservoir.
Level system is on, alarm < Level detector name > :
Check alarm or alert level sensor
level sensor cable or alert DISCONNECTED
cable connections. If message
level sensor cable has
does not clear, replace and
become disconnected
reassign module.
from module.
Level system is on, level < Level detector name > :
Disconnect and reconnect module
hardware malfunction.
CHECK MODULE
to system base. If message does
not clear, replace and reassign
module.
Alarm level sensor or
< Level detector name > :
Check alarm or alert level sensor
alert level sensor self
CHECK PROBE
connections. If message does
test failure.
not clear, replace the alarm or alert
level sensor.
Error
Level system is off,
< Level detector name > :
Disconnect and reconnect module
module hardware
Check Module
to base. If message does not
malfunction.
clear, replace and reassign module.
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...