Run a Case
9-2
Online Pumps
Access the Pumps subtab within the Setup tab and verify pump labels, display units,
and tube size. Verify safety connections and pump responses. Adjust tube sizes,
display units, and pump responses if necessary. Verify pump labels, pump direction,
and tube size on the pump local display. Verify the assignment of the primary pump
by checking the pump display for the proper names, such as “Arterial”. For Pulse
mode flow on the primary pump, use the Pulse subtab within the Pump tab to set
Rate, Width, and Base (Base only for centrifugal).
Roller Pumps
Install the tubing and set occlusion. Pumps used for venting or suction should be
fully occluded. Verify that the default direction for offline pumps matches the system
direction.
Centrifugal Pumps
Set the minimum flow alarm by using the Pumps subtab within the Setup tab. Install
the disposable pump(s) but do not start the drive motor until ready to prime the
circuit.
Power
Check the power source and battery charge level by accessing the Power
subtab within the System tab. Test the battery switching function by temporarily
disconnecting the AC (Mains) power supply. Check that the audible battery alert
tone sounds, the power status indicator LED on the front panel flashes, a message is
displayed on the CCM and is posted in the Messages subtab within the Aux tab, an
estimated available battery time is displayed above the Power Source/Battery Status
icon, and the entire system is functional. Reconnect the system to the AC (Mains)
power supply.
Modules
Verify that the module LEDs change to green for each module in the Perfusion
Screen. Check the Module Info subtab within the System tab to verify critical
information regarding the modules in the Perfusion Screen.
Gas System
(if used)
Access the Gas subtab within the Setup tab and verify the gas blend and set the
FiO2 alarm limit. Wait for the 15 minute warm up period to complete. Secure gas
outlet line to oxygenator inlet. Calibrate the oxygen analyzer after the warm up period
is complete and source gases are connected.
Occluder
(if used)
Verify that the occluder icon reads CAL. Calibrate the occluder, after placing the
tubing in the occluder head. Verify the occluder label and any safety connections to
the primary pump via the Occluder subtab within the Setup tab.
Temperature Sensors
See that the room temperature appears on each temperature sensor. Using the
Temperature subtab within the Safety tab or by touching the icon, verify temperature
labels and the temperature alarm limits for each sensor. Adjust the limits, if
necessary.
CCM
Verify that there are no question marks or red Xs on any of the device icons in the
Perfusion Screen. Verify that the correct time is displayed. Check the Message Area
and Messages subtab within the Aux tab for any potential error or status messages.
Check the Settings subtab within the System tab to verify the Perfusion Screen and
to adjust the audio tone volume.
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...