Roller Pumps
5-5
Pole Mounting
Select the desired pole position and secure the mounting arm to the pole. Select the
desired plate position and secure plate. Ensure the pump mounting lever is in the up
position. Grasp roller pump firmly with two hands. Tilt pump forward and insert the
mounting tab into the front of the mounting plate. Lower the rear of the pump onto the
plate ensuring that the mounting lever clicks into place.
If using the Descending Pump Bracket, make sure to tighten the set screw to lock the
bracket securely in position.
Note:
Ensure pole mounted pump can be hand cranked without interference from
poles or other devices. Ensure pole mounted pumps do not interfere with hand
cranking other pumps. Ensure that the Descending Pump Bracket does not impede
movement around the base.
Descending Pump Bracket
Set Screw
Dual Pumps Pole Mount
Rotating the Raceway
The roller pump raceway can be rotated
for placement of tubing to reduce tubing
length. The raceway can be positioned
± 90° from the center position in 15°
increments. To change the raceway
position, locate the spring-loaded lever on
the rear of the pump. Pull the lever away
from the pump housing.
Keeping the lever pulled, rotate the pump
raceway to the desired position. Release
the lever and rotate the pump raceway
until the lever locks into place. The
raceway should now be fixed.
Ensure the raceway is locked in place
before operating the roller pump.
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...