Centrifugal Control Unit
6-3
Motor Receptacle
Drive Motor
Select the desired pole position and secure the flexible mounting arm to the pole.
Note:
To reduce the potential for an air embolism, the drive motor is typically
positioned so that the pump will be near and just below the outlet of the reservoir
from which the blood is being pumped.
Flexible Mounting Arm
Select the desired pole position and secure the flexible mounting arm to the pole.
Set the mounting arm to the desired position:
• Turn the tension control knob counterclockwise to
loosen the joints of the mount arm. Manually support the
flexible arm when adjusting the tension control knob. Do
not loosen the knob beyond the stop.
• Position the arm as desired and tighten the tension
control knob.
Screw the drive motor onto the threaded stud. Make sure the drive motor is fully
installed on the threaded stud. Make sure the drive motor is securely mounted on the
mounting arm.
Note:
Always manually support the flexible arm when adjusting the tension control
knob.
Centrifugal Motor Mount Arm
To mount the drive motor, secure the bracket on a pole or
other vertical support and secure the drive motor to the
bracket by adding the washer and tightening the bracket
knob.
Insert the drive motor cable into the receptacle on the centrifugal drive unit. Remove
the safety cover from the drive motor.
CAUTION:
Do not sterilize the centrifugal drive motor as sterilization may
damage the motor.
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...