Central Control Monitor
4-75
Add Temperature Sensors
A maximum of 8 temperature sensors is allowed in a Perfusion Screen.
1. Select Add.
2. Select Temperature Sensor as device type.
3. Name the temperature sensor. Choose name from the list or choose
Enter New Name
and use the keyboard to enter a device name.
Set the Temperature Limit
A temperature limit for each temperature sensor can be set. An alarm condition will
be generated if the temperature exceeds the limit. The default value is 42.5 °C. The
available range for temperature limit is between 1.0 °C and 42.5 °C.
To change the alarm limit, touch the
Alarm Limit
button and use the arrows to set the
new limit. Touch the
Alarm Limit
button again to store the new value.
Choose
Next
to continue.
To Add Device:
Select Device Type:
Air Detector
Level Detector
Occluder
Pressure Sensor
Flow Sensor
Temperature Sensor
CDI 100
CDI 500
Current defined:
Maximum allowed:
0
8
Enter New Name >
Or Enter New Name:
Cancel
Set Temperature Limit:
42.5
Alarm Limit ( C)
Bladder
CPG1 Temp
CPG2 Temp
Select From List:
Select Device Name:
VEN Temp
ART Temp
ART Temp
Next >
Set Temperature Limit:
42.5
Alarm Limit ( C)
+ 1.0 °C
+ 0.1 °C
– 0.1 °C
– 1.0 °C
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...